Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG6-42
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6.2 Menu System - IntSHS Menu
Note!
Each virtual valve V1 to V8 is normally only
assigned to one gas inlet, “Gas1” to “Gas8”
.
Double assignments are possible, but only
meaningful if the unit is congured accordin-
gly.
Label Description
Off Switches the assigned component (valve or pump) off.
On Switches the assigned component (valve or pump) on.
Test Switches the assigned component (valve or pump) on and off every second.
V1 Valve at selected gas inlet is assigned the value “V1”
V2 Valve at selected gas inlet is assigned the value “V2”
V3 Valve at selected gas inlet is assigned the value “V3”
V4 Valve at selected gas inlet is assigned the value “V4”
V5 Valve at selected gas inlet is assigned the value “V5”
V6 Valve at selected gas inlet is assigned the value “V6”
V7 Valve at selected gas inlet is assigned the value “V7”
V8 Valve at selected gas inlet is assigned the value “V8”
Sample
The gas inlet is a sample gas inlet. The valve assigned to it is controlled via
the Sample software option.
Note!
There is only one option Sample . This means that in order to control two sample
gas valves during autocalibration, the value Sample must be assigned to two
inlets. During calibration, both valves are then controlled synchronously.
Pump
The pump is controlled by the Pump software option.
Note!
There is only one option
Pump. This means that in order to control two pumps
during autocalibration, the lines “Pump1” and “Pump2” must be assigned the value
Pump. During calibration, both pumps are then controlled synchronously.
Chapter 7 includes detailed descriptions
of the conguration and execution of calibra-
tions with valves.
Tab. 6-4: Parameter IntSHS options