Emerson Process Management GmbH & Co. OHG 8-19
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
8
Troubleshooting
8.4 Troubleshooting on Components
Fig. 8-1: X-STREAM X2GP
8.4.1 Opening X-STREAM Analyzers
8.4.1.1 How To Open X-STREAM X2GP
8.4.1.2 How To Open X-STREAM X2GK
Remove the top cover after loosening the 12
screws.
12 screws on top
of the instrument
Fig. 8-2: X-STREAM X2GK
If your instrument is equipped with a handle
• loosen the 6 screws at the front panel,
• to only get access to the cover screws,
push frame and handle about 2 cm / 1"
towards the rear.
Note!
To completely remove frame and handle,
you need to disconnect all gas and electrical
connections and push frame and handle over
the rear panel.
• push the cover towards the rear and re-
move it.
• remove the 4 screws for the cover,
(2 screws on each side of the instrument