6.2 Menu System - Calibration Setup
6.2.3.2 Calibration Menu
This parameter determines whether the to-
lerance check is active during calibration.
Available options:
10%: Tolerance check is active, the limit is
set to 10% (not changeable). A system
message must be manually acknow-
ledged ( 6.2.2.5, page 6-12)
AutoClr: As 10 %, except that a system mes-
sage is automatically acknowledged
after 2 or 3 minutes.
Off:
Tolerance check is deactivated
Sets calibration gas concentrations
6.2.3.2.1, page 6-24
Species analog outputs behaviour during
calibrations. Available options:
Yes:
analog outputs are “frozen” & concen-
tration alarms are surpressed
No:
the analog outputs follow the concentra-
tion. This may cause trouble if connec-
ted to a data acquisition system.
The time required when switching to zero or
span gas so that the gas lines are completely
lled with the new gas is entered here.
Possible values: 0 .. 600 seconds
If the system is set up accordingly, the access
code for level 2 must be entered to gain ac-
cess to this menu.
Note!
For further details on calibration procedures,
see chapter 7 Maintenance.
Setup..
Calibration..
Calibration gases..
Tol.Check: Aus
Hold on Cal: Yes
Purge time: 12 s
Page 1
Note!
The following lines are only available when
the “Valves” parameter in the INSTALLED
OPTIONS”menu is set to a parameter other
than none.