Software Menus
6.2 Menu System - Calibration Setup
6.2.3.2.3.1 AutoCal in Menu
Setup..
Calibration..
Interval time..
AutoCal in..
Note!
This menu appears only, if at least one interval
time has been set.
ZeroAll: 1 h
ZeroAll: 15 min
ZSCalAll: 1 h
ZSCalAll: 15 min
Note!
In case there is an interval time specied
for one procedure only (ZeroAll or ZSCalAll;
6.2.3.2.3, page 6-26), the related lines
for the other procedure will be hidden in this
menu!.
The rst two lines allow to enter a time to
elapse from the current moment, before the
next zero calibration for all channels is started.
Preset is the remaining time until the next zero
calibration starts.
Possible values: Any value up to the preset
time interval.
Lines 3 & 4 allow to enter similar data for a
combined zero & span calibration.
Possible values: Any value up to the preset
time interval.