7.4 Calibration Procedures
Type of Procedure
Menu Page
(CONTROL - )
Valves
Simulta-
neously
Calibrated
Channels
More Information
Manual calibration
ZERO CALIBRATION...
optional single channel page 7-15
SPAN CALIBRATION...
Advanced calibration
ADV.CALIBRATION - ZEROALL!
required all channels page 7-18ADV.CALIBRATION - SPANALL!
ADV.CALIBRATION - ZSCALALL
!
Remote calibration n.a. (via Modbus or Dig IN) recommended all channels page 7-29
Unattended calibration n.a. (via interval time) required all channels page 7-34
Properly conguring and performing calibrations is important to ensure proper
analyzer function. For this reason several calibration related SETUP and CONTROL
menus and their submenus can be protected by different access codes.
In the following sections this manual does not note where to enter access codes.
For information about which calibration related menus can be access code pro-
tected
Chapter 6 - CONTROL menus, page 6-5 and SETUP menus, page 6-14.