Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 6-45
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6
Software Menus
Protocol: MODB RTU
MODB-Mode: 32Bit
ID-Number: 2
Interface: RS485
6.2.3.6 Communications Settings
If the system is set up accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
Page 1
Selects Modbus protocol for the serial inter-
face.
Available options:
MODB RTU, Sensor
Note!
The Sensor option is for maintenance pur-
poses only.
Sets Modbus mode of operation.
Available options:
32Bit (=Daniel mode),
16BitLow (=Modicon mode, LOW word
rst)
16BitHi(g)h (=Modicon mode, HIGH word
rst)
Enter instrument ID for network.
Possible values: 1 .. 254
MAC ID of the Ethernet port (visible only if
"Interface" is set to Ether)
6.2 Menu System - Communications settings
Selects installed RS interface.
Available options:
RS232, RS485, Ether(net)
Setup..
Communication..
Page 2
Baud rate: 19200
Parity: None
MAC: AB234CDE56
Select baud rate of the serial interface.
Available options:
2400, 4800, 9600, 19200
Sets whether a parity bit is used.
Available options:
None, Even, Odd
Note!
To view rst 2 lines, set "Interface" to other
than Ether
Note 1
Chapter 9 includes a detailed descrip-
tion of the Modbus parameters.
A general description of Modbus is also to
be found in the appendix.
Note 2
To be able to use the optional Ethernet in-
terface, enable MODB RTU and Ether.