Emerson Process Management GmbH & Co. OHG 6-47
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6
Software Menus
6.2 Menu System - Alarm Setup Menu
Multi-channel unit:
Press the key to enter the SELECT COM-
PONENT menu to change the settings for a
different channel.
Hysteresis: 5.00 %
FlowLimit: 1 l/min
FlowLimAlm: Low FS
A hysteresis level for the switchpoints can be
entered here (as % of upper range value).
Permissible values: 0 ... 50 %.
Enter this line to specify whether an installed
ow sensor is used to only measure the ow
(Off FS) or to measure the ow and activate
alarms (Low FS).
Available options: Off FS, Low FS
Off FS: Alarm functionality is switched off,
an assigned output is set to FailSafe mode
(relay is activated).
Low FS: Alarm functionality is switched on,
output is set to FailSafe mode (alarm deac-
tivates relay). Relay is de-energized if ow
is below given limit.
Page 2
If a ow sensor is assigned to the selected
channel ( 6.2.3.5.1, page 6-44) enter this
line to specify a ow limit, to activate an alarm
as a function of ow dropping below this limit
(see next menu line).
Permissible values:
equivalent to 0 ... 2000 ml/min
Note!
The unit for ow measurement is taken from
the corresponding line in the DISPLAY SE-
TUP menu (
6.2.3.1.6, page 6-21).