Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
X-STREAM X2
Emerson Process Management GmbH & Co. OHGA-38
Instruction Manual
HASX2E-IM-EX
02/2012
A.7 Assignment of Terminals and Socket
A
.7.2 Field Housings
Pin Signal
P3.1 Input 1
P3.2 Input 2
P3.3 Input 3
P3.4 Input 4
P3.5 Input 5
P3.6 Input 6
P3.7 Input 7
P3.8 GND for Inputs 1-7
P3.9 not used
P3.10 Output 5, NC
P3.11 Output 5, NO
P3.12 Output 5, COM
P4.1 Output 6, NC
P4.2 Output 6, NO
P4.3 Output 6, COM
P4.4 Output 7, NC
P4.5 Output 7, NO
P4.6 Output 7, COM
P4.7 Output 8, NC
P4.8 Output 8, NO
P4.9 Output 8, COM
P4.10 Output 9, NC
P4.11 Output 9, NO
P4.12 Output 9, COM
P2.1 Output 10, NC
P2.2 Output 10, NO
P2.3 Output 10, COM
P2.4 Output 11, NC
P2.5 Output 11, NO
P2.6 Output 11, COM
P2.6 Output 12, NC
P2.8 Output 12, NO
P2.9 Output 12, COM
P2.10 Output 13, NC
P2.11 Output 13, NO
P2.12 Output 13, COM
Digital InputsDigital Outputs
Pin Signal
P3.1 Input 8
P3.2 Input 9
P3.3 Input 10
P3.4 Input 11
P3.5 Input 12
P3.6 Input 13
P3.7 Input 14
P3.8 GND for Inputs 8-14
P3.9 not used
P3.10 Output 14, NC
P3.11 Output 14, NO
P3.12 Output 14, COM
P4.1 Output 15, NC
P4.2 Output 15, NO
P4.3 Output 15, COM
P4.4 Output 16, NC
P4.5 Output 16, NO
P4.6 Output 16, COM
P4.7 Output 17, NC
P4.8 Output 17, NO
P4.9 Output 17, COM
P4.10 Output 18, NC
P4.11 Output 18, NO
P4.12 Output 18, COM
P2.1 Output 19, NC
P2.2 Output 19, NO
P2.3 Output 19, COM
P2.4 Output 20, NC
P2.5 Output 20, NO
P2.6 Output 20, COM
P2.6 Output 21, NC
P2.8 Output 21, NO
P2.9 Output 21, COM
P2.10 Output 22, NC
P2.11 Output 22, NO
P2.12 Output 22, COM
Digital InputsDigital Outputs
XSTA: Standard strip with standard and optional signals XSTD: First optional strip with
optional 7 Dig Inputs and 9 Dig Outputs
XSTD: Second optional strip with another set of
optional 7 Dig Inputs and 9 Dig Outputs
Power terminals
Flameproof (cast) housingFieldhousing
Note!
Line and neutral
terminals with
built-in fuse holders
L
PE
N
Power cord
entry
L= Line
N= Neutral
PE= Protective Earth
Service Port Connector -
Serial RS 232 interface
Pin
9
8
7
6
Pin
5
4
3
2
1
Ethernet connector
for Modbus
Pin 1 Pin 8
Pin no.
Signal
1 TX+
2 TX-
3 RX+
6 RX-
other not used
Pin Signal
P2.1 Channel 1, (+) 4 (0) - 20 mA
P2.2 Channel 1, GND
P2.3 Channel 2, (+) 4 (0) - 20 mA
P2.4 Channel 2, GND
P2.5 Channel 3, (+) 4 (0) - 20 mA
P2.6 Channel 3, GND
P2.7 Channel 4, (+) 4 (0) - 20 mA
P2.8 Channel 4, GND
P2.9 not used
P2.10 not used
P2.11 not used
P2.12 not used
P3.1 not used
P3.2 not used
P3.3 Output 1 (Failure), NC
P3.4 Output 1 (Failure), NO
P3.5 Output 1 (Failure), COM
P3.6 Output 2 (Maintenance Request), NC
P3.7 Output 2 (Maintenance Request), NO
P3.8 Output 2 (Maintenance Request), COM
P3.9 Output 3 (Out of Spec), NC
P3.10 Output 3 (Out of Spec), NO
P3.11 Output 3 (Out of Spec), COM
P3.12 Output 4 (Function check), NC
P4.1 Output 4 (Function check), NO
P4.2 Output 4 (Function check), COM
P4.3 not used
P4.4
P4.5
P4.6
P4.7
P4.8
P4.9
P4.10
P4.11
P4.12
Relay Outputs
**)
Analog Outputs
Serial Interface
*)
*)
See table below
**)
Conguration of relay output termi-
nals as per standard factory setting
(NAMUR status signals)
Signal Terminals Strips
Assignment of serial interface terminals
Terminal
MOD 485/
2 wire
MOD 485/
4 wire
RS 232
P4.4 SER1 Common Common Common
P4.5 SER2
not used not used
RXD
P4.6 SER3
not used not used
TXD
P4.7 SER4
not used
RXD1(+)
not used
P4.8 SER5 D1(+) TXD1(+) Common
P4.9 SER6
not used not used not used
P4.10 7
not used not used not used
P4.11 8
not used
RXD0(-)
not used
P4.12 9 D0(-) TXD0(-)
not used
Pin no.
RS 232
1 Common
2 RXD
3 TXD
4
not used
5 Common
6
not used
7
not used
8
not used
9
not used