Troubleshooting
8.2.1 Analyzer related messages
8.2.1 Analyzer Related Messages
Message
Status level
Explanation
Description Actions
FlowAlm
Check request
Explanation:
The internal ow control
detected a ow problem
The detected ow is too low
or missing due to a leak, not
limited to the instrument‘s
internal gas path
Check the external and inter-
nal gas path for leakage and
plugging
If applicable check internal
pump function
LocalAccess
Function check
Explanation:
This status message
is activated when a
parameter is changed
using the front panel
keys
Someone has changed a
parameter by using the front
panel keys
Press the HOME key: The
status message is reset when
returning to the measurement
screen
Acknowledge the message
via menu CONTROL.. -
ACKNOWLEDGEMENTS..
Note!
This resets ALL status mes-
sages!
NotSampleGas
Function check
Explanation:
The gas currently owing
is not the expected
sample gas
Other than sample valve is
opened within installed valve
block
Switch on sample valve
Installed pump is switched off Switch on the pump
Instrument is currently in cali-
bration mode
Wait for calibration to end
After calibration the sample
gas valve has opened, but the
post-purge time has not yet
elapsed
Wait for purge time to elapse.
If appropriate reduce purge
time
Pressure
Out of spec
Explanation:
The internal pressure
control detected a
pressure problem
The detected pressure is too
low or missing due to a leak
inside or outside the instru-
ment
Check the external and inter-
nal gas path for leakage and
plugging
If applicable check internal
pump function