Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG8-4
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
Message
Status level
Explanation
Description Actions
NVRAM fail
Check request
Explanation:
NVRAM test failed
Installed NVRAM and/or
electronics board defective Replace
ROMmemory
Failure
Explanation:
Wrong code checksum
Installed FLASH memory
defective Replace board
Simulation
Function check
Explanation:
This message does not
appear during operation
modes!
The instrument is set into a
debugging mode by service
personnel
Switch analyzer off and on
again to exit the debugging
mode
Switch off the related simu-
lation parameters (in service
level or by Modbus command)
Warm-up
Function check
Explanation:
This message may
require temperature
monitoring enabled
within the service level
The warm-up time has not yet
elapsed after last analyzer
restart
Wait for warm-up time to
elapse
Temperature of analyzer
components or physical ther-
mostatted compartment is not
within the congured range
Wait for instrument to heat up
check internal heater for pro-
per function
SensTimeout
Failure
Explanation:
No internal communi-
cation to XSP board
Check internal wiring
Check COM parameters via
SVC port
SensCmdFail
Failure
Explanation:
Sending a command to
XSP board failed
Bandwidth, currently available
for data exchange with XSP
board, too small
Acknowledge the message. If
recurring, call for service.
8.2.1 Analyzer related messages