Princeton 4411-0139 Home Security System User Manual


 
60 PI-MAX
®
4 System Manual Issue 4
3. From the host computer, click on the Focus button or select Acquisition —>
Focus
from the pull-down menu. Data acquisition will begin and an acquired image
should be displayed on the computer monitor.
If an image is successfully acquired and displayed, it has been confirmed that
the PI-MAX4 can acquire images.
If the image is out of focus, reposition the target and/or rotate the lens and then
click on
Acquire to determine the degree to which the focus has been changed.
If an image is not displayed, verify the software configuration settings and then
try again to acquire a test image.
See Figure 4-13 for a sample image.
Figure 4-13: Sample Acquired Test Image
4.
Once it has been confirmed that the PI-MAX4 is able to acquire images, turn the
PI-MAX4 I.I.T. switch OFF and close WinX.
NOTE:
If the PI-MAX4 will not be used with the lens, replace the C-
mount lens with the screw-in dust cover provided with the C-
mount adapter.
If a different adapter, will be used, switch adapters at this point.
Be sure to use the adapter’s dust cover whenever the PI-MAX4
is not being used.
4.6.5 Focusing
REFERENCES:
For complete information about focusing the PI-MAX4 camera
when using C-, F-, or spectroscopy mounts, refer to:
Section E.4, Focusing C- and F-Mount Lenses;
Section F.3, Spectrograph-Camera Rotational
Alignment and Focusing
4411-0137_0016