Chapter 14 PI-MAX4: EM Family 219
14.2.6 Configuring Kinetics
Perform the following procedure to configure a PI-MAX4: EM camera for Kinetics mode:
1. Click on the Readout expander and select Kinetics within the Readout Mode field.
2. Specify the desired Kinetics Window Height in number of rows.
NOTE:
The number of Frames per Readout is automatically
calculated and displayed in the Readout expander.
See Figure 14-14.
Figure 14-14:Configuring Readout Mode
NOTE:
When the Kinetics Window Height value is changed, the
Frames per Readout value is recalculated. Consequently, the
value specified in
Common Acquisition Settings —>
Number of Frames
must be manually updated to reflect this
new Frames per Readout value.
NOTE:
An Experiment Warning is raised indicating a mismatch in
configuration parameters.
4411-0137_0125