Chapter 13 PI-MAX4: 1024i-RF and RF Modulation 193
Example
The following Phase Sweep values have been configured:
Range Start: -720°
Range End: +360°
Increment: 90°
The total sweep length is 1080° in 90° increments/steps which corresponds to 13
data points. Since Range Start < Range End, the sweep proceeds in a positive/
counterclockwise direction.
LightField “shifts” the programmed sweep starting point so that it starts at 0°.
Table 13-1 shows the corresponding/calculated values for each step in the sweep.
Table 13-1: Phase Sweep Information for Sweep > 360°
Phase
Table Step
Phase Table
Value
Programmed
Phase Angle
Cumulative
Sweep Total
Increment
1
0
° -720° [Range Start] 0° 90°
2
90
° -630° 90° 90°
3
180
° -540° 180° 90°
4
270
° -450° 270° 90°
5
0
° -360° 360° 90°
6
90
° -270° 450° 90°
7
180
° -180° 540° 90°
8
270
° -90° 630° 90°
9
0
° 0° 720° 90°
10
90
° 90° 810° 90°
11
180
° 180° 900° 90°
12
270
° 270° 990° 90°
13
0
° 360° [Range End] 1080° 90°