Princeton 4411-0139 Home Security System User Manual


 
110 PI-MAX
®
4 System Manual Issue 4
c. The SyncMASTER ON button enables the SyncMASTER trigger output from the
SyncMASTER1 and SyncMASTER2 connectors on the
AUX I/O cable. The
frequency for the SyncMASTER outputs may then be configured, as well as the
AUX Output signal at the AUX OUT connector on the rear of the PI-MAX4.
When SyncMASTER is enabled, the output of the SyncMASTER1
connector will be at the frequency specified on the Internal Trigger
Frequency field.
The output of the SyncMASTER2 connector will be at the same frequency
as that of
SyncMASTER1. However, a delay can be specified so the edges
of
SyncMASTER2 will occur after the edges of SyncMASTER1.
When using the AUX Output signal from the SuperSYNCHRO to trigger a
piece of equipment, enter the Auxiliary pulse delay time. Enter the pulse
width needed to trigger the equipment.
Figure 6-14 illustrates SuperSYNCHRO Timing with SyncMASTER on.
Figure 6-14: Typical SuperSYNCHRO Timing: SyncMASTER On
4411-0137_0079