222 PI-MAX
®
4 System Manual Issue 4
14.2.8.1 Configuring the Software Parameters:
NOTE:
The following procedure is based on LightField. When using
different application software, modifications to this procedure
may be required. Basic familiarity with the LightField software
is assumed. If this is not the case, review the software manual
or have it available while performing this procedure.
Configure the software parameters as follows:
• Readout expander
— Readout Mode: Available modes are read from the camera. Select Kinetics.
— Kinetics Window Height: Enter the desired number of exposed rows.
— Frames per Readout: This value is automatically calculated based on the
specified window height.
• Common Acquisition Settings expander
— Number of Frames: Enter the Frames per Readout value calculated by the
system.
— Exposures per Frame: 1
• Region of Interest expander: Use the functions on this expander to define the
region of interest (ROI.)
— Clicking on Full Sensor selects the full window (without binning)
•Sensor expander
— Temperature Setpoint: The default temperature setting is read from the
camera. When the array temperature reaches the set temperature, that status will
be reported as
Locked in the Sensor expander and in the Status bar.
• Trigger expander
— Trigger Response
Valid values are:
• Readout Per Trigger
• Shift Per Trigger
— Trigger Source
Valid values are:
• Internal
• External
REFERENCES:
Refer to the LightField User’s Manual, document number
4411-0125, for information about configuring external trigger
sources.