138 PI-MAX
®
4 System Manual Issue 4
e. Click on the Trigger Out tab to configure the following parameters:
• SyncMASTER Trigger Out
When Enabled (i.e., checked,) the frequency of the SyncMASTER1 output
is the same as the frequency of the
Trigger In signal. Refer to step 13.b on
page 135.
• SyncMASTER2 Delay from SyncMASTER1: measured in µsec.
The output of the
SyncMASTER2 signal is identical to that of the
SyncMASTER1 output. However, a delay, in microseconds, may be
configured so that the leading and trailing edges
SyncMASTER2 will occur
a specified time AFTER the corresponding edges of
SyncMASTER1.
• Auxiliary Parameters
When using the
Aux Out signal to trigger an external piece of equipment,
configure the following parameters:
— Width
Specifies the width of the desired
Aux Out trigger pulse. This will vary
depending on the specific equipment being controlled.
— Delay
Specifies the delay between the start of
SyncMASTER1 and the start of
the
Aux Out trigger pulse.
See Figure 7-19.
Figure 7-19: Typical SuperSYNCHRO Dialog: Trigger Out Tab
f.
Click OK to configure all SuperSYNCHRO parameters and dismiss the dialog.
4411-0137_0059