4 System Manual Issue 4
G.3 Quick Start Guide
When a camera system is shipped with a spectroscopy-mount adapter kit, a quick start guide
for mounting a PI-MAX4 with spectroscopy-mount adapter to an Acton spectrograph is
included. Typically, the guide assumes that there is a spectroscopy-mount adapter
pre-mounted to the PI-MAX4 camera.
G.4 Standard C-, F-, and Spectroscopy-Mount Adapters
Figure G-2 shows three different mounts that are available for use on a PI-MAX4 camera.
Figure G-2: Standard PI-MAX4 Mount Adapters
Perform the following procedure to replace one adapter with another:
1. Use the supplied screwdriver to remove the four (4) Phillips head screws that secure the
current adapter to the front of the PI-MAX4.
NOTE:
If the Phillips end of the screwdriver is not available, pull the
driver shaft out of the handle, flip the shaft, and insert the flat
end into the handle.
2. Remove the existing adapter and store it in a safe place.
3. Using the four (4) screws removed in step 1, mount the new adapter to the front of the
PI-MAX4.
C-Mount F-MountSpec-Mount
Mounting Screws
(4 Places)
4411-0137_0175