Barco R9843501 Pet Fence User Manual


 
15. Troubleshooting
1. Check the USB connection between LEX unit and MCU
- Plug and reconnect the USB connector of the LEX unit from/to the MCU
- On the AutoAlignment Head, do a power cycle
2. Try to nd out if the USB port on the MCU is failing
- Plug in the USB connector of the LEX unit into another USB port on the MCU
- On the AutoAlignment Head, do a power cycle
3. Check the RJ45 connections of the SFTP Cat5e cable
- Unplug and reconnect both RJ45 connectors of the SFTP Cat.5e cable between the AutoAlignment Head and the LEX unit
15.4 Camera
Since the laser pointer is installed next to the spectrometer, there is a small parallax erro r between the laser
pointer and the spect rometer Field of View, depending on the distance to these devices.
Camera settings
To guarantee proper working, the camera settings should never be changed! The default settings of the various selection switches
are listed below:
Power switch: ON
Rotary dial: M
Focus selection switch on the camera lens: MF
Image 15-6
Camera settings
No image from the camera
No Live view from c urrent se lected cam era is visible on the Position page of the XDS RACU user interface
Possible cause Action
Lens cap on the Camera lens Remove the lens cap
R59770509 AUTOALIGNMENT HEAD GEN II 26/11/2013 157