14. Replacing devices
3. Enter the serial ID of the new camera.
Note: Serial number of the camera can be read from the X DS RACU S erver Console or from the label attached to the bottom
of the camera.
Serial num ber of the camera can also be read from the Select the cam era to be replaced drop down list (image 14-3).
XDS RACU automatically assigns an ID to each connected cam era. The ID has format cameraX, where X is a digit that
is increased by one for eve ry newly conn ected camer a. The last connected camera has the highest d igit in its ID.
4. Press the Replace button to conrm the replacement.
Image 14-4
Conrming the camera replacement
A message appearing in the feedback section shows the progress information.
14.2 Replacing a spectrometer
How to replace a Spectrometer?
1. In the O ptions page, navigate to Replace Device > Spectrometer.
Image 14-5
Section: Spectrometer
2. Select the spectrometer to be replaced from the drop down list.
Image 14-6
Selecting old spectrometer
3. Enter the serial ID of the new spectrometer.
Note: Serial number of the spec trometer can be read from t
he XDS RACU Server Console or from the label attached to the
device.
Serial number of the spectrometer can be also read from the Select the spectrometer to be replaced dr op down list (im-
age 14-6). XDS RACU automatically assigns an ID to each con nected spectrometer. The ID has format spect rom eterX,
where X is a digit that is increased by one for e very newly connected spectrometer. The last connected spectrometer
has the highest digit in its ID.
4. Press the Replace button to conrm the replacement.
Image 14-7
Conrming the spectrometer replacement
A message appearing in the feedback
section shows the progress information.
14.3 Replacing a laser
How to replace a Laser?
1. In the Options page, navigate to Replace Device > Laser.