276 Daniel MeterLink fails to connect or crashes using Windows® 2000
Section 7: Maintenance and troubleshooting Reference, Installation, and Operations Manual
June 2013 3-9000-743 Rev S
3. Select the Trust access to Visual Basic Project checkbox and click the OKAY button to
apply your selections.
Figure 7-11 Excel® Trusted Access Setting
7.4.4 Daniel MeterLink fails to connect or crashes using Win-
dows® 2000
Daniel MeterLink fails to connect or crashes on a Windows® 2000 machine with Service Pack 4
when attempting a direct serial or modem connection with a Series 100 Option BoardDaniel
3410 Series Ultrasonic Gas Flow Meter.
Windows® 2000 Service Pack 4 has a problem that will prevent Daniel MeterLink from
connecting or cause Daniel MeterLink to crash when attempting a direct serial or modem
connection to a Daniel 3400 Series Gas Ultrasonic Flow Meter. This is a Known Issue documented
in the Microsoft® Knowledge Base Article 824301 “Script Error”- Error Message When You Try
to Connect to a Server by Using a Remote Access Connection. A Hotfix is available from
Microsoft®. See below, to find out where to find this patch.
Windows 2000 Excel® Menu Windows XP Excel® Menu