150 Temperature and Pressure page
Section 5: Startup and meter configuration Reference, Installation, and Operations Manual
June 2013 3-9000-743 Rev S
1. To configure the Live Temperature, plus associated alarms, choose the settings listed in
the table below):
Table 5-16 Live Temperature Inputs
Daniel MeterLink
Display Name
Options and Guidelines
Live temperature, Min input
• Enter the temperature (°C or °F) that corresponds to a 4 mA input signal.
The temperature must be within the range [-273.15 °C, 200 °C].
Live temperature, Max input
• Enter the temperature (°C or °F) that corresponds to a 20 mA input
signal. The temperature must be within the range [-273.15 °C, 200 °C].
Temperature alarm, Low limit
• Enter a value (°C or °F) within the range [-130 °C, 200 °C]. An alarm is
generated when the temperature is at or below this limit value.
Temperature alarm, High limit
• Enter a value (°C or °F) within the range [-130 °C, 200 °C]. An alarm is
generated when the temperature is at or above this limit value.
Alarm Selection
• The Alarm selection determines what value to use while a live input is in
alarm. This value is common with the pressure Alarm selection so when
one is changed, the other will change to match.
• Choose Last good value or Fixed value