Frequency test mode
The Mark III meter provides a mode of operation for individually testing the frequency output
signal pairs. Entering, configuring, and exiting this mode is accomplished by setting data points
using the Daniel MeterLink
Tools - Outputs Test screen as described in the table below. Note that
when a frequency output pair is in the test mode, the frequency output is considered to be
invalid since it does not represent the selected frequency content.
The maximum length of time that a frequency output pair can remain in the test mode is
specified via the NonNormalModeTimeout data point. Note that this data point applies to both
frequency output pairs (as well as various other tests). The NonNormalModeTimeout data
point can be changed using the Daniel MeterLink
Edit/Compare Configuration screen. It can be
set within the range [1, 30 min] and has a default value of 2 min.
Table 6-12 Data points for frequency outputs test
Daniel MeterLink
Display Name
Data Points, Options and Guidelines
Frequency output 1
Output setting
or
Frequency output 2
Output setting
Data points affected:
Freq1TestModeOutputPercent
Freq2TestModeOutputPercent
Options:
Enter the integer percentage of full scale for the corresponding frequency output pair
within the range [0, 150 %] where full scale is as specified via the corresponding
Freq1MaxFrequency or Freq2MaxFrequency data point.
Guidelines:
The specified test value takes effect within one batch period of clicking the Start button
(see below).
For options and guidelines, see timeout discussion below and Daniel MeterLink online help.
Frequency output 1
Start(Stop)
or
Frequency output 2
Start(Stop)
Data points affected:
IsFreq1EnableTest
IsFreq2EnableTest
Options:
Click on the Start button to enter the test mode (TRUE) (Note: the Start button will become
the Stop button after clicking on it.)
Click on the Stop button to exit the test mode (FALSE) (Note: the Stop button will become
the Start button after clicking on it.)
Guidelines:
The specified test value takes effect within one batch period of clicking the Start button
(see below).
For further information, see timeout discussion below and Daniel MeterLink online help.