Teledyne API Model T360/T360M Operation Manual Operating Instructions
81
These default Data Channels can be used as they are, or they can be customized
from the front panel to fit a specific application. They can also be deleted to
make room for custom user-programmed Data Channels.
Appendix A-5 lists the firmware-specific DAS configuration in plain-text format.
This text file can either be loaded into APICOM and then modified and uploaded
to the instrument or can be copied and pasted into a terminal program to be sent
to the analyzer.
NOTE
Sending a DAS configuration to the analyzer through its COM ports will replace the existing
configuration and will delete all stored data. Back up any existing data and the DAS configuration
before uploading new settings.
4.7.2.1. DAS CONFIGURATION LIMITS
The number of DAS objects are limited by the instrument’s finite storage
capacity. For information regarding the maximum number of channels,
parameters, and records and how to calculate the file size for each data channel,
refer to the DAS manual downloadable from the TAPI website at
http://www.teledyne-api.com/manuals/.
4.7.2.2. VIEWING DAS DATA AND SETTINGS
DAS data and settings can be viewed on the front panel through the following
touchscreen buttons sequence.
07272B DCN6552