3.8. INITIAL CALIBRATION PROCEDURE
The next task is to calibrate the analyzer.
To perform the following calibration you must have sources for zero air and span
gas available for input into the sample port on the back of the analyzer. See
Section 3.6 for instructions for connecting these gas sources.
While it is possible to
perform
this procedure with any range setting we
recommend that you perform this initial checkout using the 500 ppm range.
The following procedure assumes that the instrument does not have any of
the available Zero/Span Valve Options installed. See Section 5.4 for
instructions for calibrating instrumen
ts with Z/S valve options.
1. Set the Analog Output Range:
SETUP X.X
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SETUP X.X RANGE CONTROL MENU
MODE
SET UNIT EXIT
Press this button to set
the analyzer for
SNGL
DUAL or AUTO ranges
SETUP X.X RANGE: 500.000 CONC
0 0 5 0 0 .0 ENTR EXIT
To change the value of the
reporting range span, enter the
number by pressing the key under
each digit until the expected value
appears.
Press this button to select the
concentration units of measure:
PPB, PPM, UGM, MGM
EXIT ignores the new setting and
returns to the RANGE CONTROL
MENU.
ENTR accepts the new setting and
returns to the
RANGE CONTROL MENU.
SETUP X.X RANGE: 500.000 Con
c
0 0 0 5 0 .0 ENTR EXIT
07272B DCN6552