Teledyne T360M Carbon Monoxide Alarm User Manual


 
Teledyne API - T360/T360M, 360E/360EM Appendix A Menu Trees (05233C DCN6552)
A-14
Setup Variable Numeric
Units
Default
Value
Value Range Description
CO_CONST2 — 0.2110
0.356
20,22,23
0.367
15
1.458
9,12
1.4625
18
1.448
4
0.192
8
0.187
3
0.1196
24
0–10 CO calculation constant.
ET_MEAS_GAIN — 1 0.0001–9.9999
Electrical test gain factor for
measure reading.
ET_REF_GAIN — 1 0.0001–9.9999
Electrical test gain factor for
reference reading.
ET_TARGET_DET mV 4375 0–5000
Target detector reading during
electrical test.
ET_TARGET_CONC PPM 40,
400
3, 8
1–9999.99
Target concentration during
electrical test.
ET_CONC_RANGE Conc. 50,
5000
3, 8
0.1–50000
D/A concentration range during
electrical test.
STD_TEMP ºK 321 1–500
Standard temperature for
temperature compensation.
STD_PRESS "Hg 28.5,
28.7
8
,
28.8
12, 18
,
28.1
4
1–50
Standard pressure for pressure
compensation.
48 BENCH_SET ºC
Warnings:
43–53
0–100
Optical bench temperature set
point and warning limits.
68,
62
19,23
WHEEL_SET ºC
Warnings:
63–73,
57–67
19,23
0–100
Wheel temperature set point and
warning limits.
50 O2_CELL_SET
14
ºC
Warnings:
45–55
30–70
O
2
sensor cell temperature set
point and warning limits.
STD_O2_CELL_TEMP
14
ºK 323 1–500
Standard O
2
cell temperature for
temperature compensation.
ZERO_APPLY_IN_CAL
5
ON OFF, ON
ON applies auto-reference offset
and dilution factor during
zero/span calibration;
OFF disables both.
(Only applicable if
ZERO_ENABLE is ON.)
07272B DCN6552