4.13.3. Analog Output Step Test
This test can be used to check the accuracy and proper operation of the analog
outputs. The test forces all four analog output channels to produce signals
ranging from 0% to 100% of the full scale range in 20% increments. This test is
useful to verify the operation of the data logging/recording devices attached to the
analyzer.
To begin the Analog Output Step Test press:
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
Exit-Exit
returns to the
DIAG menu
DIAG
A
NALOG OUTPUT
PREV NEXT ENTR EXIT
DIAG AOUT ANALOG OUTPUT
0% EXIT
Performs
analog output
step test.
0% - 100%
DIAG AOUT ANALOG OUTPUT
[0%] EXIT
Pressing the “0%” button while performing the test will
pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same button again
will resume the test.
SETUP X.X ENTER DIAG PASS: 818
8 1 8 ENTR EXIT