Teledyne T360M Carbon Monoxide Alarm User Manual


 
Teledyne API - T360/T360M, 360E/360EM Appendix A Menu Trees (05233C DCN6552)
A-12
Setup Variable Numeric
Units
Default
Value
Value Range Description
CO_DWELL Seconds 0.2 0.1–30
Dwell time before taking
measure or reference sample.
CO_SAMPLE Samples 1 1–30
Number of samples to take in
measure or reference mode.
PRE_FILT_SIZE
5, 19
Samples 16 1–50 Moving average pre-filter size.
FILT_SIZE Samples 750,
720
9, 12
200
3, 8
1000
19, 23
1–1000 Moving average filter size.
FILT_ASIZE Samples 48,
20
3, 8
,
40
20, 22
1–1000
Moving average filter size in
adaptive mode.
FILT_DELTA PPM 4,
0.7
5
15
3, 8
0.15
9, 12
0.4
19, 23
0.2
20, 22
1–1000
Absolute change to trigger
adaptive filter.
FILT_PCT % 10
5
20, 22, 23
1–100
Percent change to trigger
adaptive filter.
FILT_DELAY Seconds 90,
72
20, 22
0–180
Delay before leaving adaptive
filter mode.
FILT_ADAPT — ON ON, OFF
ON enables adaptive filter; OFF
disables it.
CO2_DWELL
10
Seconds 0.1 0.1–30
Dwell time before taking each
sample.
CO2_FILT_ADAPT
10
ON ON, OFF
ON enables CO
2
adaptive filter;
OFF disables it.
CO2_FILT_SIZE
10
Samples 48 1–300 CO
2
moving average filter size.
CO2_FILT_ASIZE
10
Samples 12 1–300
CO
2
moving average filter size in
adaptive mode.
CO2_FILT_DELTA
10
% 2 0.01–10
Absolute CO
2
conc. change to
trigger adaptive filter.
CO2_FILT_PCT
10
% 10 0.1–100
Percent CO
2
conc. change to
trigger adaptive filter.
CO2_FILT_DELAY
10
Seconds 90 0–300
Delay before leaving CO
2
adaptive filter mode.
CO2_DIL_FACTOR
10
1 0.1–1000
Dilution factor for CO
2
. Used only
if is dilution enabled with
FACTORY_OPT variable.
O2_DWELL
14
Seconds 1 0.1–30
Dwell time before taking each
sample.
O2_FILT_ADAPT
14
— ON ON, OFF
ON enables O
2
adaptive filter;
OFF disables it.
O2_FILT_SIZE
14
Samples 60 1–500
O
2
moving average filter size in
normal mode.
O2_FILT_ASIZE
14
Samples 10 1–500
O
2
moving average filter size in
adaptive mode.
07272B DCN6552