Teledyne API - T360/T360M, 360E/360EM Appendix A Menu Trees (05233C DCN6552)
A-15
Setup Variable Numeric
Units
Default
Value
Value Range Description
ZERO_DWELL
3, 5, 8
Seconds,
Minutes
5
7,
3
5
1–60,
1–30
5
Dwell time after closing or
opening zero scrubber valve.
ZERO_SAMPLES
3, 5, 8
Samples 15,
750
5
,
1000
19
1–1000
Number of zero samples to
average.
ZERO_FILT_SIZE
3, 5, 8
Samples 5,
1
5
1–100
Auto-zero offset moving average
filter size.
ZERO_LIMIT
3, 5, 8
Ratio 1.2,
1.15
3, 8
,
1
5
0–5
Minimum auto-zero ratio allowed;
must be greater than this value
to be valid.
ZERO_CAL
3, 5, 8
Ratio 1.18 0.5–5 Calibrated auto-zero ratio.
CO_TARG_ZERO1 Conc. 0
-100.00–
999.99
Target CO concentration during
zero offset calibration of range 1.
CO_TARG_MID1_1 Conc. 50
5
,
300
0.01–9999.99
Target CO concentration during
mid-point #1 calibration of range
1.
CO_TARG_MID2_1 Conc. 50
5
,
300
0.01–9999.99
Target CO concentration during
mid-point #2 calibration of range
1.
CO_SPAN1 Conc. 40,
400
3, 8
0.01–9999.99
Target CO concentration during
internal span calibration of range
1.
CO_SLOPE1 — 1 0.001–999.999 CO slope for range 1.
CO_OFFSET1 — 0 -10–10 CO offset for range 1.
CAL_BOX_TEMP1 ºC 30 0–100
Calibrated box temperature for
range 1.
CO_TARG_ZERO2 Conc. 0
-100.00–
999.99
Target CO concentration during
zero offset calibration of range 2.
CO_TARG_MID1_2 Conc. 50
5
,
300
0.01–9999.99
Target CO concentration during
mid-point #1 calibration of range
2.
CO_TARG_MID2_2 Conc. 50
5
,
300
0.01–9999.99
Target CO concentration during
mid-point #2 calibration of range
2.
CO_SPAN2 Conc. 40,
400
3, 8
0.01–9999.99
Target CO concentration during
internal span calibration of range
2.
CO_SLOPE2 — 1 0.001–999.999 CO slope for range 2.
CO_OFFSET2 — 0 -10–10 CO offset for range 2.
CAL_BOX_TEMP2 ºC 30 0–100
Calibrated box temperature for
range 2.
CO2_TARG_MID1_CONC
10
% 6,
800
16
0.1–1000,
0.1–2000
16
Target CO
2
concentration during
mid-point #1 calibration.
CO2_TARG_MID2_CONC
10
% 6,
800
16
0.1–1000,
0.1–2000
16
Target CO
2
concentration during
mid-point #2 calibration.
CO2_TARG_SPAN_CON
C
10
% 12 0.1–1000,
0.1–2000
16
Target CO
2
concentration during
span calibration.
07272B DCN6552