4.8.3. Single Range Mode (SNGL)
This is the default reporting range mode for the analyzer. In single range mode
both A1 and A2 are set to the same reporting range. This reporting range can be
any value between 50 ppm and 2 000 ppm.
While the two outputs always have the same reporting range, the span, signal
offset and scaling of their electronic signals may be configured for differently
(e.g., A1 = 0-10 V; A2 = 0-0.1 V). See Section 4.13.4 for instructions on
adjusting thes
e parameters.
To select SNGL range mode and to set the upper limit of the range, press:
SETUP C.3
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP C.3 RANGE MODE: SNGL
SNGL
DUAL AUTO ENTR EXIT
EXIT x 2 returns
to the main
SAMPLE display
SETUP C.3 RANGE MODE: SNGL
SNGL
DUAL AUTO ENTR EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP C.3 RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SAMPLE* RANGE = 500.000 PPM CO2 X.XXX
< TST TST > CAL SETUP