5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following
settings:
B
aud Rate: 115200
COM2 Mode Settings:
Quiet Mode ON
Computer Mode ON
MODBUS RTU OFF
MODBUS ASCII OFF
E,8,1 MODE OFF
E,7,1 MODE OFF
RS-485 MODE OFF
SECURITY MODE OFF
MULTIDROP MODE OFF
ENABLE MODEM OFF
ERROR CHECKING ON
XON/XOFF HANDSHAKE OFF
HARDWARE HANDSHAKE OFF
HARDWARE FIFO ON
COMMAND PROMPT OFF
6. Next, configure your communications software, such as APIcom. Use the
COM port determined in Step 4 and the baud rate set in Step 5. The figures
below show how these parameters would be configured in the Instrument
Properties window in APIcom when configuring a new instrument. See the
APIcom manual (PN 07463) for more details.
Note that USB configuration requires that baud rate (Section 4.11.10) of
instrument and PC match. Also, using the USB port disallows use of the rear
panel COM2 port except for multidrop communications (Section 4.11.8).