Teledyne T360M Carbon Monoxide Alarm User Manual


 
Teledyne API - T360/T360M, 360E/360EM Appendix A Menu Trees (05233C DCN6552)
A-26
Name
1
Message Text Description
COSLOPE SLOPE=1.000
CO SLOPE=1.000
7
CO slope for current range, computed
during zero/span calibration.
COSLOPE1 SLOPE1=1.000
CO SLOPE1=1.000
7
CO slope for range #1 in dual range
mode, computed during zero/span
calibration.
COSLOPE2 SLOPE2=1.000
CO SLOPE2=1.000
7
CO slope for range #2 in dual range
mode, computed during zero/span
calibration.
COOFFSET OFFSET=0.000
CO OFFSET=0.000
7
CO offset for current range, computed
during zero/span calibration.
COOFFSET1 OFFSET1=0.000
CO OFFSET1=0.000
7
CO offset for range #1 in dual range
mode, computed during zero/span
calibration.
COOFFSET2 OFFSET2=0.000
CO OFFSET2=0.000
7
CO offset for range #2 in dual range
mode, computed during zero/span
calibration.
CO2SLOPE
7
CO2 SLOPE=1.000
CO
2
slope, computed during zero/span
calibration.
CO2OFFSET
7
CO2 OFFSET=0.000
CO
2
offset, computed during zero/span
calibration.
O2SLOPE
10
O2 SLOPE=0.980
O
2
slope, computed during zero/span
calibration.
O2OFFSET
10
O2 OFFSET=1.79 %
O
2
offset, computed during zero/span
calibration.
CO CO=17.7 PPM
3
CO concentration for current range.
CO2
7
CO2=15.0 % CO
2
concentration.
O2
10
O2=0.00 % O
2
concentration.
TESTCHAN TEST=1751.4 MV
Value output to TEST_OUTPUT analog
output, selected with TEST_CHAN_ID
variable.
XIN1
12
AIN1=37.15 EU
External analog input 1 value in
engineering units.
XIN2
12
AIN2=37.15 EU
External analog input 2 value in
engineering units.
XIN3
12
AIN3=37.15 EU
External analog input 3 value in
engineering units.
XIN4
12
AIN4=37.15 EU
External analog input 4 value in
engineering units.
XIN5
12
AIN5=37.15 EU
External analog input 5 value in
engineering units.
XIN6
12
AIN6=37.15 EU
External analog input 6 value in
engineering units.
XIN7
12
AIN7=37.15 EU
External analog input 7 value in
engineering units.
XIN8
12
AIN8=37.15 EU
External analog input 8 value in
engineering units.
07272B DCN6552