Teledyne T360M Carbon Monoxide Alarm User Manual


 
Teledyne API - T360/T360M, 360E/360EM Appendix A Menu Trees (05233C DCN6552)
A-18
Setup Variable Numeric
Units
Default
Value
Value Range Description
TEST_CHAN_ID — NONE
0
NONE,
CO
MEASURE,
CO
REFERENC
E,
VACUUM
PRESSURE,
SAMPLE
PRESSURE,
SAMPLE
FLOW,
SAMPLE
TEMP,
BENCH
TEMP,
WHEEL
TEMP,
O2 CELL
TEMP
14
,
CHASSIS
TEMP,
PHT DRIVE,
TEMP4
5
Diagnostic analog output ID.
REMOTE_CAL_MODE — LOW
0
LOW,
HIGH,
CO2
10
,
O2
14
CO range or other gas to
calibrate during contact closure
or Hessen calibration.
PASS_ENABLE — OFF ON, OFF
ON enables passwords; OFF
disables them.
STABIL_FREQ Seconds 10
120
19, 23
1–300
Stability measurement sampling
frequency.
STABIL_SAMPLES Samples 25 2–40
Number of samples in
concentration stability reading.
2500 PHOTO_TEMP_SET mV
Warnings:
250–4750
0–5000
Photometer temperature warning
limits. Set point is not used.
29.92 SAMP_PRESS_SET In-Hg
Warnings:
15–32
0–100
Sample pressure warning limits.
Set point is not used.
07272B DCN6552