HP (Hewlett-Packard) HP 8360 Vacuum Cleaner User Manual


 
Note
U, V, and W-band power sensors are not available from
Hewlett-Packard. For these frequencies use the Anritsu ML83A
Power Meter with the MP715-004 (40 to 60
GHz),
the MP716A
(50 to 75
GHz),
or the MP81B (75 to 110 GHz) power sensors. Since
the Anritsu model ML83A Power Meter is not capable of internally
storing power sensor cal factors, you must manually correct the
data entry. Refer to example 2 for information on manual entry of
correction data.
5.
6.
7.
8.
9.
10.
11.
12.
Setup Synthesizer Parameters
Turn on the synthesizer and press
[PRESET).
The following occurs:
n The source module’s frequency span is displayed on the
synthesizer.
n The synthesizer’s leveling mode is automatically changed
internal to “module leveling”.
n The source module’s maximum specified power is set and
displayed.
from
Press FREQUENCY (START)
@
@
0
@
(GHz,
m@
@
IGHz).
The frequency sweep is set from 26.5 to 40 GHz.
Press
(POWER
LEVEL)
0
@.
The source module power is set to
+7
dBm
for maximum power to the device under test.
Access User Flatness Correction Menu
Press POWER
CE).
Select Fltness Menu.
Select Delete Menu Delete All. This step insures that the
flatness array is empty.
Press
(PRIOR).
Leave the delete menu and return to the previous
softkey
menu.
Select Enter Freq
(!?J
@
0
@
(GHz),
to enter 26.5 GHz as
the first correction frequency. Enter 31, 32.5, and 40 GHz to
complete the list. Notice that the frequencies are arbitrarily
spaced.
Enter Correction Data into Array
Select Mtr
Meas
Menu Measure
Corr
All. The power meter
is now under synthesizer control and is performing the sequence
of steps necessary to generate the correction information at each
frequency point.
If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the synthesizer is
Getting Started Advanced 1-41