HP (Hewlett-Packard) HP 8360 Vacuum Cleaner User Manual


 
w
OPTION NOT INSTALLED. The language selected and the
corresponding firmware/hardware necessary to run that language
is not present in the synthesizer. See “INSTALLATION” for
information on language selection.
DISPLAY IS NOT RESPONDING: Can appear on the front panel
emulator if the internal processor can not communicate with the
display properly. This error indicates a display failure or a display
connector problem.
DIVIDE BY ZERO EXCEPTION: This can only be caused by an
internal processor error. Refer to the “OPERATOR’S CHECK”
chapter for instructions on contacting a qualified service technician.
EEROM FAILED, LOST CAL: This error indicates that the
synthesizer has lost its calibration constants and may not meet
specifications. Refer to the “OPERATOR’S CHECK” chapter and
follow the local operator’s check procedures. If you are a qualified
service technician and this failure occurs, read the Calibration
Constants section in the Service Guide.
EEROM Failed !!: This error will only occur if the service
adjustment menu is accessed. Specifically, an attempt has been made
to write to a test patch and EEROM failed to store the data.
ERROR: CALIBRATION FAILED !!: This error will only occur if
the service adjustment menu is accessed. Specifically, an
Al4
sweep
ramp calibration has been attempted and failed. Run the sweep
ramp selftest, refer to “MENUS and SELFTESTS” in the Assembly
Level Repair Manual.
ERROR
-
Must first enter correction freq: This error occurs when a
correction point does not have its corresponding frequency entered
first. Refer to “Creating and Applying the User Flatness Correction
Array,” in chapter 1 of this handbook.
ERROR: Must first enter a List Frequency !!: This error occurs
when a dwell or offset value does not have its corresponding
frequency entered first. Refer to “Creating and Using a Frequency
List,” in chapter 1 of this handbook.
ERROR: Power Search Failed
!!: This error occurs when the ALC is
in the ALC search mode and is unable to level to the desired power
level. Refer to the “OPERATOR’S CHECK” chapter and follow the
local operator’s check procedures.
ERROR: Start must be < Stop !!:
This error occurs in association
with the frequency list, auto fill, feature. If the start frequency
entered is greater than the stop frequency, you will see this error.
Correct by entering a start frequency less than the stop frequency.
ERROR: Stop must be > Start
!!: This error occurs in association
with the frequency list, auto fill, feature. If the stop frequency
entered is less than the stop frequency you will see this error. Correct
by entering a stop frequency greater than the start frequency.
2a-2
Error Messages