Initiates a calibration of the external flatness. Depends on
value of CALibrat ion: PMETer : RANGe.
Initiates a calibration of the power meter flatness.
Depends on value of CALibrat ion : PMETer : RANGe.
Initiates a calibration of the source module flatness.
Depends on value of CALibrat ion : PMETer : RANGe.
l CALibration:PMETer:FLATness:NEXT?
<num>[lvlsuffix]
The parameter is the measured power that is currently produced by
the synthesizer. You must supply this parameter after measuring the
power using an external power meter. The query response is issued
after the synthesizer processes the supplied parameter and settles on
the next point to be measured. The query response is:
>o
The frequency [in Hz] that is currently produced.
0
The calibration is complete.
<O
An error has occurred and the calibration is aborted.
a
CALibration:SPAN:AUTO ONlOFFlllO
l CALibration:SPAN:AUTU?
Sets and queries the automatic sweep span calibration.
ON
OFF
A calibration is done whenever the sweep span is changed.
A calibration is done only when
CALibrat ion : SPAN
[ :
EXECuteI
is sent.
After
*RST,
the setting is OFF.
l
CALibrat ion : SPAN
[
:
EXECut
el
Causes a sweep span calibration.
l CALibrat ion :
TRACk
Causes an automatic tracking calibration procedure.
l
CORRection:ARRay[il
(<num>
[DBI
)1601*1601
l
CORRect ion :
ARRay
[il?
Sets and queries the entire 1601-point array of correction values
that can be added to the internal flatness correction array. The
1601-points are added to the internal flatness array synchronized
on the trigger output
1~s
pulses.
These TTL-level pulses are 1601
evenly spaced points across an analog sweep, or at each point in step
or list mode. Entering this array causes the
CORRection:SOURce
Cil
command to set to
ARRay.
There is one array for the foreground
state (i=O) and one for the background state (i=l). If the [i] is not
specified, the default value is i=O. After
*RST,
these arrays are
cleared.
l
CORRect ion : FLATness (<num>
[f
req suffix] ,
<num>
CDBI
32*801
l CORRection:FLATness?
S-30
Operating and Programming Reference