This
softkey
accesses the doubler amp mode softkeys. These softkeys
are applicable to instrument models with a doubler installed. The
doubler has an integral amplifier whose operation is controlled by the
instrument firmware. Its use depends on the frequency of operation
and on the calibration constants set at the factory. The instrument
defaults after preset to this automatic mode of operation which is
the specified operation.
Softkeys in this menu will allow you to turn
the doubler amplifier always on or always off. These two modes are
unspecified operation for instruments with a doubler installed. These
softkeys have no effect on instruments without a doubler.
Doubler Amp Mode AUTO
Sets the doubler amp mode to
AUTO. This is the default after
preset and must be used for specified
performance.
Doubler Amp Mode On
Turns the doubler amplifier on
regardless of the frequency of
operation. Using this mode results in
unspecified performance.
Doubler
&rip
Mode Off
Turns the doubler amplifier off
regardless of the frequency of
operation. Using this mode results in
unspecified performance.
Programming Codes
SCPI: NONE
Analyzer: NONE
See Also
Softkeys listed above.