HP (Hewlett-Packard) HP 8360 Vacuum Cleaner User Manual


 
l
ABORt
Causes the sweep in progress to abort and reset. If
INIT:
CONT is
ON it immediately restarts the sweep. The pending operation flag
(driving
*OPC,
*WAI, and *OPC?)
undergoes a transition once the
sweep is reset.
l
AM[:DEPTh]
Qnun>[PCTlIMAXimumlMINimuml~num>DB
l
AM[:DEPThl? [MAXimumlMINimum]
Sets and queries the percentage of AM when the SOURce is
INTernal.
If <num> is received with units of
dB,
the value is
converted to percent by the equation:
[%]
= 100
*
(1
-
lO**(-dB/20))
Valid ranges of dB are 0 to 40 dB. After
“RST,
the value is 50%.
l
AM:INTernal:FREQuency
Qnun>[Cfreqsuffix>]lMAXimumlMINimum
l
AM:INTernal:FREQuency? [MAXimumIMINimum]
Sets and queries the frequency (in Hz) of the internal AM source.
After
*RST,
the value is 1 kHz.
l
AM:INTernal:FUNCtion
SINusoidlSqUarelTRIanglelRAMPlNOISe
l
AM:INTernal:FUNCtion?
Sets and queries the waveform of the internal AM source. After
“RST,
the value is SINusoid.
l
AM:SOURce
INTernallEXTernal
l
AM:SOURce?
Sets and queries the source of the AM modulating signal. After
*RST,
the value is
EXTernal.
l
AM:MODE
DEEPlNORMal
a
AM:MODE?
Controls the AM depth limits of the synthesizer. The
NORMal
position is selected at
*RST.
l AM:STATe
ONlOFFlllO
l
AM:STATe?
Sets and queries the status of the AM modulation. After
*RST,
the
setting is OFF.
l
AM:TYPE
LINearlEXPonential
a
AM:TYPE?
Sets and queries the type of AM modulation. After
*RST,
the
setting is
LINear.
l
CALibration:AM:AUTO
ONlOFFlllO
a
CALibration:AM:AUTO?
Sets and queries the automatic modulator calibration switch.
S-28
Operating and Programming Reference
HP 8360
User’s Handbook