HP (Hewlett-Packard) HP 8360 Vacuum Cleaner User Manual


 
SCPI
COMMAND SUMMARY
HP 8360
User’s Handbook
is not affected by *RST. When you change the value from ON to
OFF, everything except calibration data is initialized or destroyed.
In particular, data in instrument state and all save/recall registers
are destroyed.
l
SYSTem:VERSion?
This query returns a formatted numeric value corresponding to
the SCPI version number to which the synthesizer complies. The
response has the form YYYY.V where the Ys represent the year
version (i.e. 1990)and the V represents an approved revision number
for that year. This is a query only and therefore does not have an
associated *RST state.
l
TRIGger[:IMMediate]
Causes the trigger event to occur regardless of other settings in the
subsystem. This event does not affect any other settings in this
subsystem.
This command has no effect unless the synthesizer is in the wait for
trig
state. If the synthesizer is in the wait
for
trig
state, it performs
its trigger action. This is an event and has no “RST condition.
l
TRIGger:ODELay
<num>[time
suffix]
IMAXimumlMINimum
l
TRIGger:ODELay.
7
[MAXimumlMINimum]
Sets and queries the trigger output delay, the time between when
the source is settled, (when Bit 1 of the Standard Operation Status
Register makes a negative transition), and the trigger out signal is
sent.
l
TRIGger:SOURce
IMMediatelBUSIEXTernal
l
TRIGger:SOURce?
Sets and queries the source of the trigger event.
a
TSWeep
This is a convenience command that does the equivalent of
ABORt;INITiate[:IMMediate]
l
UNIT:AM
DBIPCT
l
UNIT:AM?
Sets and queries the default units for AM depth.
The *RST value is PCT.
l
UNIT:POWer [lvl suffix]
l
UNIT:POWer?
Sets and queries the default power subsystem units.
*RST
value is DBM.
Operating and Programming Reference
S-55