command to set the switch will cause ROSC:SOUR:AUTO OFF to
be done also. The *RST value is automatically determined.
l
ROSCillator:SOURce:AUTO
ONlOFFlllO
a
ROSCillator:SOURce:AUTO?
Sets and queries the automatic reference selection switch.
The *RST value is 1.
l
STATus:OPERation:CONDition?
Queries the Standard Operation Condition register.
l
STATus:OPERation:ENABle
<mua>
l
STATus:OPERation:ENABle?
Sets and queries the Standard Operation Enable register.
The
STATus:PRESet
value is 0.
l
STATus:OPERation[:EVENt]?
Queries the Standard Operation Event Register.
This is a destructive
read.
l
STATus:OPERation:NTRansition
Cnum>
l
STATus:OPERation:NTRansition?
Sets and queries the Standard Operation Negative Transition Filter.
The
STATus:PRESet
value is 0.
l
STATus:OPERation:PTRansition
<rum>
l
STATus:OPERation:PTRansition?
Sets and queries the Standard Operation Positive Transition Filter.
After
STATus:PRESet,
all used bits are set, to
1’s.
l
STATUS:PRESet
This command presets the following enable and transition registers:
MSIB,
OPERation,
QUEStionable,
and
SRECeiver.
ENABle
Is set to all
0s.
NTRansition
Is set to all
0s.
PTRansition
All bits used are set to
1s.
Unused bits remain
OS.
l
STATus:QUEStionable:CONDition?
Queries the Data Questionable Condition Register.
l
STATus:QUEStionable:ENABle
<mm>
l
STATus:QUEStionable:ENABle?
Sets and queries the Data Questionable SRQ Enable register.
The
STATus:PRESet value is 0.
S-48
Operating and Programming Reference
HP 8360