HP (Hewlett-Packard) HP 8360 Vacuum Cleaner User Manual


 
HP 8360
User’s Handbook
SCPI
COMMAND SUMMARY
l
SWEep:MANual:POINt
<num>lMAXimumlMINimum
l
SWEep:MANual:POINt?
[MAXimumlMINimum]
Sets and queries the step point number to go to and lock. The value
is a
unitless
value that is limited between 1 and the number of points
requested. This command has no effect on the instrument unless:
the sweep mode is set to manual and the sweep generation is set to
stepped mode.
*RST value is 1.
l
SWEep : MANual
[
:
RELat
ive]
<mm>
l
SWEep : MANual
[
:
RELat
ive] ?
Sets and queries a percent of sweep to go to and lock. This command
has no effect unless: the sweep mode is set to manual and the sweep
generation is set to analog. *RST value is 0.50.
l
SWEep:MARKer:STATe
ONlOFFlllO
l
SWEep:MARKer:STATe?
Sets and queries the state of marker sweep. When ON, the frequency
sweep limits are taken to be the positions of marker 1 and marker 2.
*RST
value is 0.
l
SWEep :
MARKer
: XFER
This transfers the values of marker 1 and marker 2 frequencies into
start and stop frequency.
l
SWEep:MODE
AUTOlMANual
l
SWEep : MODE?
Selects and queries the manual sweep mode switch.
AUTO
The sweep is under the control of the
INIT
and
SWEEP subsystems.
MANual
FREQ:MANual, SWEep:MANual[:RELative], and
SWEep:MANual:POINt control the output.
*RST value is AUTO.
l
SWEep:POINts
<num>lMAXimumlMINimum
a
SWEep :
POINts? [MAXimum
I
MINimum]
Sets and queries the number of points in a step sweep. When points
is changed, SWEep : STEP is modified by the equation
STEP
=
SPAN/POINTS.
Span is normally an independent variable but is changed to STEP
x POINTS if both of these parameters are changed in the same
message.
*RST value is 11.
l
SWEep : STEP
Cnum>
[f
req suffix]
I
MAXimum
I
MINimum
l
SWEep:STEP.
7
[MAXimum I MINimum]
Sets and queries the size of each frequency step. : STEP is governed
by the equation
STEP
=
SPAN/POINTS.
Operating and Programming Reference
S-51