entry point> a string response that identifies the
paragraph number in the
HP 8360 Assembly-Level Repair Manual
to
begin the troubleshooting procedure.
l
DIAGnostics:TEST:CONTinue
Causes the
selftest
execution to continue when paused for raw data
examination. Does nothing in other conditions.
l
DIAGnostics:TEST:DATA:DESC?
Returns the description string of the raw data examined during a
selftest. It returns
in other conditions.
l
DIAGnostics:TEST:DATA:MAXimum?
Returns the upper test limit for the raw data point examined.
Returns “0” in other conditions.
l
DIAGnostics:TEST:DATA:MINimum?
Returns the lower test limit for the raw data point examined.
Returns “0” in other conditions.
l
DIAGnostics:TEST:DATA:VALue?
Returns the raw data value for the raw data point examined.
Returns a “0” in other conditions.
l
DIAGnostics:TEST:DISable
(<num>)l*?IALL
Prevents the listed selftests from being selected. If ALL is sent then
all of the selftests are disabled. *RST causes
DIAG:TEST:ENAB
ALL to
execute.
l
DIAGnostics:TEST:ENABle
CCnum>3l*?lALL
Enables the listed selftests to execute. If ALL is sent then all of the
selftests are enabled.
*RST
causes DIAG:TEST:ENAB ALL to execute.
l
DIAGnostics:TEST[:EXECute]
<num>
The specified
selftest
is executed. Normal instrument operation is
suspended and the instrument state is restored upon exiting the
selftest
mode.
l
DIAGnostics:TEST:LOG:SOURce
ALLIFAIL
l
DIAGnostics:TEST:LOG:SOURce?
Sets and queries the source for the raw data logging. ALL specifies
that all raw data points are displayed. FAIL selects only those data
points out of the test limits. Both commands are executable in
selftest
mode. After
*RST,
the setting is FAIL.
Operating and Programming Reference