FREQUENCY
2
This
softkey
sets the frequency step size in the swept frequency step
mode. The step size may be set from 1 Hz to 10 GHz. The factory
preset step size is 100 MHz. Step size values are entered using the
entry area.
If an underline cursor appears under a digit in the entry display, then
the value will be modified by the up/down arrow keys or the rotary
knob. The increment/decrement size in this case is the underlined
digit by the power of 10.
If the up/down function is on (asterisk next to key label) and the
cursor is not under one of the active entry area digits, then frequency
value is changed by the up/down size using either the up/down arrow
keys or the rotary knob.
SCPI: FREQuency:STEP[:INCR]
<num>[freq
suffix] or
MAXimumlMINimum
Analyzer: SF or SHCF <num> [HzlKzlMzlGz]