Function Group
SWEEP
Menu Map
7
Description
This softkey lets you set a sweep time for frequency sweeps or power
sweeps. The sweep time range is 10 ms to
2OOs,
but the fastest sweep
time is constrained by the frequency span. The fastest possible sweep
can be determined automatically:
1. Press SWEEP (MENU), this reveals the sweep menu keys.
2.
Select more
If3
, to scroll to the next page of the sweep menu.
3.
Select
SwpTime
Auto,
to set the sweep time to automatic. The
synthesizer calculates the fastest possible calibrated sweep time for
any sweep span.
Whenever you press
(SWEEPTIME),
the active entry area displays
the current sweep time and whether the sweep time is coupled to
the frequency span (far right hand side displays:
AUTO). If the word
AUTO is not displayed then the sweep time auto function is off.
Programming Codes
SCPI: SWEep[:FREQ
uency]:TIME
<num>[time suffix] or
MAXimum]
MINimum
Analyzer: ST <num>
[sclms]
See Also
Power Sweep
“Power Level and Sweep Time Operation,” in Chapter 1.
“Programming Typical Measurements,” in Chapter 1.
SwpTime
Auto
Function Group
SWEEP
Menu Map
7
S-78
Operating and Programming Reference
HP 8360
User’s Handbook