HP (Hewlett-Packard) HP 8360 Vacuum Cleaner User Manual


 
Internal Pulse Mode Gate
Programming Codes
SCPI: PULM:INTernal:WIDTh <num>[time suffix]]MAXimum]MINimum
Analyzer: NONE
See Also
0,
1
MOD
a so see “Pulse” and “Modulation”,
Internal Pulse Mode Auto
Function Group
(MOD)
Menu Map
4
Description
This
softkey
(Option 002 only) is the default mode of generating
internal pulses. It is not synchronized to any trigger signal. An
asterisk next to the key label indicated that this mode is selected.
Programming Codes
SCPI: PULM:INTernal:TRIGger:SOURce
INTernal
Analyzer: NONE
See Also
0,
1
MOD
a so see “Pulse” and “Modulation”.
Internal Pulse Mode Gate
Function Group
(MOq)
Menu Map
4
Description
This
softkey
(Option 002 only) logically “ANDs” the internal pulse
generator with a gating signal supplied from an external source.
Programming Codes
SCPI:
PULM:INTernal:GATE ON(OFF(l]O
PULM:INTernal:TRIGger:SOURce
INTernal
Analyzer: NONE
See Also
(MOD), also see “Pulse” and “Modulation”.
HP 8360
User’s Handbook
Operating and Programming Reference
I-1 1