HP (Hewlett-Packard) HP 8360 Vacuum Cleaner User Manual


 
Power Offset
Function Group
POWER
Menu Map
5
Description
This softkey changes the mapping of absolute power parameters on
input to the synthesizer. It does not change the RF output produced
by the synthesizer. The equation used to determine the displayed
value is:
Entered or Displayed Power = Hardware Power (ALC)
+
Active
Offset.
Programming Codes
SCPI:
POWer:OFFset:STATe ON]OFF]llO
POWer:OFFset
<num>[DB]]MAXimum(MINimum(UP(DOWN
Analyzer: NONE
See
Also
[POWER LEVEL) and POWER
(FV1ENU).
Power Slope
Function Group
POWER
Menu Map
5
Description
This softkey lets you compensate for system, cable, and waveguide
variations due to changes in frequency, by linearly increasing or
decreasing power output as the frequency increases. RF slope values
may range from -2.50 to
i-2.50
dB per GHz. The power at the
beginning of the sweep equals the current power level. An asterisk
next to the key label indicates that this feature is active.
Programming Codes
SCPI:
POWer:SLOPe:STATe ON(OFFIl]O
POWer:SLOPe
<num>[DB](MAXimum]MINimum]UP]DOWN
Analyzer:
SLl
function on, SLO function off. Note that because SL
functions in the fundamental units of
dB/Hz,
you program the SL
code
SLmdt,
where m is 1 (on) or 0 (off); d is the numerical value
P-6
Operating and Programming Reference
HP 8360
User’s Handbook