Teledyne 360E Carbon Monoxide Alarm User Manual


 
Operating Instructions Model 360E Instruction Manual
The M360E solves this problem by allowing the user to select a scaled reporting range for the analog outputs
that only includes that portion of the physical range relevant to the specific application. Only the reporting range
of the analog outputs is scaled, the physical range of the analyzer and the readings displayed on the front panel
remain unaltered.
6.8.2. Reporting Range Modes
The M360E provides three analog output range modes to choose from.
Single range (SNGL) mode sets a single maximum range for the analog output. If single range is
selected (See Section 6.78.3) both outputs are slaved together and will represent the same
measurement span (e.g. 0-50 ppm), however their electronic signal levels may be configured for
different ranges (e.g. 0-10 VDC vs. 0-.1 VDC – See Section 6.9.4.1).
Dual range (DUAL) allows the A1 and A2 outputs to be configured with different measurement spans
(See Section 6.8.4) as well as separate electronic signal levels (See Section 6.9.4.1).
Auto range (AUTO) mode gives the analyzer to ability to output data via a low range and high range.
When this mode is selected (See Section 6.8.5) the M360E will automatically switch between the two
ranges dynamically as the concentration value fluctuates.
Range status is also output via the external digital I/O status outputs (See Section 6.15.1.1).
To select the Analog Output Range Type press:
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE EXIT
SETUP X.X RANGE CONTROL MENU
MODE SET UNIT EXIT
SETUP X.X RANGE MODE: SNGL
SNGL DUAL AUTO ENTR EXIT
EXIT Returns
to the Main
SAMPLE Display
Go To
Section
6.7.3
Go To
Section
6.7.4
Go To
Section
6.7.5
Only one of the
range modes may
be active at any
time.
78 05232 Rev B3