Teledyne 360E Carbon Monoxide Alarm User Manual


 
Calibration Procedures Model 360E Instruction Manual
7.5. Manual Calibration Checks with Zero/Span Valves
Zero and span checks using the VARIOUS zero/span valve options available for the M360E are similar to that
described in Section 7.3, except that the zero and calibration operations are initiated directly and independently
with dedicated keys CALZ and CALS.
To perform a manual calibration check of an analyzer with a valve option installed, use the following method.
STEP ONE: Connect the sources of Zero Air and Span Gas as shown in Figures 7-3 through 7-6.
STEP TWO: Perform the zero/span check.
Wait until STABIL
falls below 1.0
ppm.
This may take
several minutes.
The value of STABIL
may jump
significantly. Wait
until STABIL falls
below 1.0 ppm. This
may take several
minutes.
SPAN CAL M STABIL=XXX.X PPM CO2 =X.XXX
< TST TST > ZERO SPAN CONC EXIT
SAMPLE STABIL=XXX.X PPM CO2 =X.XXX
< TST TST > CAL CALZ CALS SETUP
ZERO CAL M STABIL=XXX.X PPM CO2 =X.XXX
< TST TST > ZERO CONC EXIT
SAMPLE STABIL=XXX.X PPM CO2 =X.XXX
< TST TST > CAL CALZ CALS SETUP
EXIT returns to the main
SAMPLE dis
p
la
y
ACTION:
Record the
CO2 readings
presented in the
upper right corner of
the display.
ACTION:
Record the
CO
2
readings
presented in the
upper right corner of
the display.
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL CALZ CALS SETUP
Scroll to the STABIL test
function.
146 05232 Rev B3