Teledyne 360E Carbon Monoxide Alarm User Manual


 
Operating Instructions Model 360E Instruction Manual
NOTE
While Hessen Protocol Mode can be activated independently for COM1 and COM2, The
TYPE selection affects both Ports.
6.15.4.5. Setting The Hessen Protocol Response Mode
The Teledyne Instruments implementation of Hessen Protocol allows the user to choose one of several different
modes of response for the analyzer.
Table 6-30: Teledyne Instruments Hessen Protocol Response Modes
MODE ID MODE DESCRIPTION
CMD
This is the Default Setting. Reponses from the instrument are encoded as the traditional command format.
Style and format of responses depend on exact coding of the initiating command.
BCC
Responses from the instrument are always delimited with <STX> (at the beginning of the response, <ETX>
(at the end of the response followed by a 2 digit Block Check Code (checksum), regardless of the command
encoding.
TEXT
Responses from the instrument are always delimited with <CR> at the beginning and the end of the string,
regardless of the command encoding.
To Select a Hessen response mode, press:
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SETUP X.X HESSEN RESPONSE MODE :CMD
<SET SET> EDIT EXIT
SETUP X.X COMMUNICATIONS MENU
ID HESN COM1 COM2 EXIT
SETUP X.X HESSEN VARIATION: TYPE 1
SET> EDIT EXIT
SETUP X.X HESSEN RESPONSE MODE :CMD
BCC TEXT EDIT ENTR EXIT
Press to
change
response
mode.
ENTR key accepts the
new settings
EXIT key ignores the new
settings
6.15.4.6. Hessen Protocol Gas ID
The Model 360E Analyzer is a single gas instrument that measures CO
2
. As such it’s default gas ID has already
been set to 310. There is no need to change this setting.
134 05232 Rev B3