Teledyne 360E Carbon Monoxide Alarm User Manual


 
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES
4. If gas flows through the instrument when it is disconnected from its sources of zero air, span gas or
sample gas, the flow problem is most likely not internal to the analyzer. Check to make sure that:
All calibrators/generators are turned on and working correctly.
Gas bottles are not empty or low.
Valves, regulators and gas lines are not clogged or dirty.
11.2.2.2. Low Flow
1. Check if the pump diaphragm is in good condition. If not, rebuild the pump (See Section 9.3.2). Check
the Spare Parts List for information of pump rebuild kits.
2. Check for leaks as described in Section 9.3.3. Repair the leaking fitting, line or valve and re-check.
3. Check for the sample filter and the orifice filter for dirt. Replace filters (See Sections 9.3.1 and 11.5.1
respectively).
4. Check for partially plugged pneumatic lines, orifices, or valves. Clean or replace them.
5. If an IZS option is installed in the instrument, press CALZ and CALS. If the flow increases then suspect
a bad sample/cal valve.
11.2.2.3. High Flow
The most common cause of high flow is a leak in the sample flow control assembly or between there and the
pump. If no leaks or loose connections are found in the fittings or the gas line between the orifice and the pump,
rebuild/clean the sample flow control assembly as described in Section 11.6.1.
11.2.2.4. Displayed Flow = “XXXX”
This warning means that there is inadequate gas flow. There are four conditions that might cause this:
1. A leak upstream or downstream of the flow sensor
2. A flow obstruction upstream or downstream of the flow sensor
3. Bad Flow Sensor Board
4. Bad pump
To determine which case it is, view the sample pressure and sample flow functions on the front panel. If the
sample pressure is reading abnormally low, then the cause is likely a flow obstruction upstream of the flow
sensor. First, check the sample filter and make sure it is not plugged and then systematically check all the other
components upstream of the orifice to ensure that they are not obstructed.
If the sample pressure is reading normal but the sample flow is reading low then it is likely that the pump
diaphragm is worn or there is an obstruction downstream of the flow sensor.
11.2.2.5. Actual Flow Does Not Match Displayed Flow
If the actual flow measured does not match the displayed flow, but is within the limits of 720-880 cc/min, adjust
the calibration of the flow measurement as described in Section 6.9.8.
05232 Rev B3 203