APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
LIN_RATIO2 — 1 0.01–100
Measure/reference ratio measured
during linearity adjustment for range
2.
LIN_CORRECT2 — 1 0.001–999.999 Linearity correction factor for range 2.
LIN_TARGET_CONC Conc 300 1–10000
Target concentration during linearity
adjustment.
LIN_NORM_CONC PPM 300 0.01–10000
Target concentration during linearity
adjustment normalized for T/P.
LIN_RATIO — 1 0.01–100
Measure/reference ratio measured
during linearity adjustment.
LIN_CORRECT — 1 0.001–999.999 Linearity correction factor.
CO2_COMP_ENABLE — OFF ON, OFF
ON enables CO
2
compensation; OFF
disables it.
CO2_COMP_CONC % 0 0–20
CO
2
concentration to compensate
for.
CO_CONST1 — 500 100–50000 CO calculation constant.
CO_CONST2 — 1.448 0–10 CO calculation constant.
ET_MEAS_GAIN — 1 0.0001–9.9999
Electrical test gain factor for measure
reading.
ET_REF_GAIN — 1 0.0001–9.9999
Electrical test gain factor for
reference reading.
ET_TARGET_DET mV 4375 0–5000
Target detector reading during
electrical test.
ET_TARGET_CONC PPM 40 1–10000
Target concentration during electrical
test.
ET_CONC_RANGE Conc. 50 0.1–50000
D/A concentration range during
electrical test.
STD_TEMP ºK 321 1–500
Standard temperature for
temperature compensation.
STD_PRESS "Hg 28.1 1–50
Standard pressure for pressure
compensation.
BENCH_SET ºC
48
Warnings:
43–53
0–100
Optical bench temperature set point
and warning limits.
WHEEL_SET ºC
68
Warnings:
63–73
0–100
Wheel temperature set point and
warning limits.
O2_CELL_SET
5
ºC
50
Warnings:
45–55
30–70
O
2
sensor cell temperature set point
and warning limits.
STD_O2_CELL_TEMP
5
ºK 323 1–500
Standard O
2
cell temperature for
temperature compensation.
CO_SPAN1 Conc. 40 1–10000
Target CO concentration during span
calibration of range 1.
CO_SLOPE1 — 1 0.001–999.999 CO slope for range 1.
CO_OFFSET1 — 0 -10–10 CO offset for range 1.
CO_SPAN2 Conc. 40 1–10000
Target CO concentration during span
calibration of range 2.
CO_SLOPE2 — 1 0.001–999.999 CO slope for range 2.
CO_OFFSET2 — 0 -10–10 CO offset for range 2.
O2_TARG_SPAN_CONC
5
% 20.95 0.1–100
Target O
2
concentration during span
calibration.
A-18 05233 Rev G.4